Wait about 1 minute and select On to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Needle Extend Denied
To prevent the collision of autosampler components, the system did not execute
the command to extend the autosampler needle.
Autosampler - Needle Failed to Extend, Reset Autosampler
The autosampler needle failed to extend properly.
Possible Cause
1
Critical
Sampler
Error.
2
40 PSI line is
out of range.
3
Needle is
bent.
4
Needle sensor
failed.
Autosampler - Needle Failed to Retract, Reset Autosampler
The autosampler needle failed to retract properly.
Possible Cause
1
Critical
Sampler
Status Line messages
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
will reset.
c
If the problem persists, select
touchpad.
d
Wait about 1 minute and select
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
a
Check pressure and adjust if necessary.
b
Check for pinched air lines supplying needle cylinder.
c
On the analyzer touchpad, select
d
Wait about 1 minute and select On to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
Replace needle.
Call Siemens Service for assistance.
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
on the analyzer
Off
to restart the
On
Off
.