Siemens ADVIA 2120 Operator's Manual page 320

Table of Contents

Advertisement

%MICRO/
%HYPO
RATIO
Parameter Key
Dilution Factor
The dilution factor for the RBC channel is 83.33333E-6.
RBC
Red blood cell count
MCV
Mean Corpuscular Volume
RDW
Red Cell Volume Distribution Width
CHCM
Corpuscular Hemoglobin Concentration Mean
CH
Cellular Hemoglobin Content
HDW
Hemoglobin Concentration Distribution Width
HCT
Hematocrit
Number of red cells
(RBC R Count x RBC Valid Cells) ÷ (RBC R Count + RBC P Count)
RBC R Count
RBC Valid Cells
The number of valid electronic pulse signals detected from flowcell events.
Methods
%MICRO ÷ %HYPO
Number of cells in the RBC
area (1) of the RBC Scatter
Cytogram.
8-51