Main Power Input Line Filter; Removing The Main Power Input Line Filter; Installing The Main Power Input Line Filter - Siemens SIRECUST 1280 Service Manual

Bedside monitors
Table of Contents

Advertisement

E331.E2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
dures
of
Section
4.9.5.
(Omit
Steps 3 and 6 in Section 4.9.5.)
Reinstall
the
Curve
Board
fol-
lowing
procedures
outlined
in
- Section 4.4.3. (Omit Step 6).
Complete the procedure of Sec-
tion 4.11.2 to reinstall the Z-Amp
Board
reassemble
the
monitor.
(Omit Step 6 in Section 4.10.2.)
Main Power Input Line Filter
Refer to Figure 4-15.
Removing the Main Power
input Line Filter
Remove
the
outer
case
of
the
monitor
following
the
procedure
outlined in Section 4.1.1.
Remove the Power Supply Board
following the procedure outlined in
Section 4.2.1.
3.
Remove and save the two Phillips-
head
screws
and
washers
(А)`
securing the line filter housing to
the monitor chassis.
Remove
and
save
the
chassis
grounding
post (B), lock-washer,
and nut on the filter housing.
Lift the Power Supply Line Filter
assembly out of the monitor.
Installing the Main Power
Input Line Filter
Position the Main Power Line Filter
unit so that the mounting holes in
the line filter housing
align with
the mounting holes in the chassis,
and the filter grounding tab aligns
with the chassis post hole.
Insert
Phillips-head
screws
with
washers
(A)
into
the
mounting
holes,
and
tighten the screws
to
secure the housing to the chassis.
Figure 4-15.
Main Power Input Line Filter
Page 4 - 25