Siemens ADVIA 2120 Operator's Manual page 206

Table of Contents

Advertisement

2
Faulty barcode
reader.
3
Window of
barcode reader
is dirty, or line
of sight is
obstructed.
4
Rack barcode
label
improperly
placed,
obscured, or
absent.
Autosampler - Rack Scan Failed
Barcode reader is unable to scan.
Possible Cause
1
Faulty barcode
reader cabling.
2
Faulty barcode
reader.
Autosampler - Shuttle Gate Open
Input queue cannot advance the rack because the shuttle gate is not closed.
Possible Cause
1
Shuttle gate
only partially
closed.
2
Faulty sensor
or sensor
wiring.
Status Line messages
See 1a, b, c
a
Clean barcode reader window.
b
Make sure there is no visual obstruction between the
barcode reader window and the rack.
c
If problem persists, call Siemens Service for
assistance.
Make sure the barcode label is correctly placed on the
rack, is clean, and meets specifications.
Corrective Action
a
At the analyzer touchpad, select
b
Wait about 1 minute and select
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
See 1a, b, c
Corrective Action
a
Check for an obstruction to the shuttle gate.
b
Push the shuttle gate to check for spring resistance.
CAUTION
Call Siemens Service for assistance. Only Siemens
Service personnel are authorized to take this corrective
action:
Check sensor and associated wiring.
Off
.
On
to restart the
7-37