Autosampler - Startup Failed, Reset Autosampler
The autosampler has failed to start up. A reset is required.
Corrective Action
1
Open the autosampler access door.
2
Close the autosampler access door. The autosampler resets.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Tube Too High
Tube is sitting too high in rack, or tube exceeds specifications, halting
autosampler.
Possible Cause
1
Tube is
improperly
positioned.
2
Tube is too
long.
Autosampler Startup Complete
The autosampler has successfully completed the startup process.
Autosampler - Unexpected Car Position, Reset Autosampler
The autosampler car is not in the proper position. A reset is required.
Corrective Action
1
Open the autosampler access door.
2
Close the autosampler access door. The autosampler resets.
3
If the problem persists, call Siemens Service for assistance.
Awaiting Header
Software communication error.
Possible Cause
7-38
Corrective Action
a
Push tube down.
b
At the analyzer touchpad, select
continue operations.
a
Remove tube from rack and aspirate manually.
b
At the analyzer touchpad, select
continue operations.
Corrective Action
to
Start/Stop
Start/Stop
to
Status Line messages