Checking and adapting the line voltage
Fig. 1

Checking the nominal line voltage

Measure the line voltage with the digital multimeter on-site at the wall outlet.
Record the line voltage.

Adaptation of the nominal line voltage

NOTE
At the bottom of the control console loosen the hand screw and the protective conductor
cable shoe (accessible through the cassette compartment).
Loosen the corrugated hose from the holding clamps.
Lift the control console at the handle recesses and set it down on a table or similar base.
Loosen the 4 Allen screws (1/Fig. 1, wrench size 4mm) and lift the cover upwards.
Loosen the cheese-head screw (2/Fig. 2) and the 2 nuts (3/Fig. 2) and hinge the control
unit to the side.
At the terminal block of the transformer T1 change the conductor connection from T1.6
(220 V) to T1.7 (240 V) (4/Fig. 2).
Then close the control console in the reverse order, set it on the mobile stand, fasten it
with the hand screw and connect the protective conductor cable shoe.