Siemens ADVIA 2120 Operator's Manual page 250

Table of Contents

Advertisement

RBC Irregular Flow Rate - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
RBC Irregular Flow Rate - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
RBC Reagent Empty - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The reagent is empty, and the autosampler has
halted. The specified stop criterion is user definable in the Customize Menu,
System Setup, Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The stop counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation. Select Start/Stop to continue the run.
RBC Reagent Expired
Present date is beyond the RBC reagent expiration date, as determined by
encoded reagent container barcode label or as calculated from installation date
and open container stability.
Corrective Action
Replenish expired reagents and update the Reagent Installation window. Go to
Logs, Reagent Log, Reagent Installation.
RBC Reagent Low - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message, indicating that the reagent level is low. The
specified alarm criterion is user definable in the Customize Menu, System Setup,
Alarm/Stop Criteria, Reagent Conditions.
Status Line messages
7-81

Advertisement

Table of Contents
loading

This manual is also suitable for:

Advia 2120i

Table of Contents