Autosampler - Calibration Required
To start the analyzer, initial calibration values for the Car Index position, Mixer
Shuttle position and the Mixer Aspirate position must be stored in flash memory.
Call Siemens Service for assistance. Only Siemens Service personnel are
authorized to take these corrective actions.
Possible Cause
1
Autosampler
CPU board
contains new,
unprogrammed
flash memory.
2
Flash memory
has lost the
stored values.
Autosampler - Car Alignment Error, Reset Autosampler
The autosampler car is out of alignment. A reset is required.
Corrective Action
1
Open the autosampler access door.
2
Close the autosampler access door. The autosampler resets.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Car Initialization Failed, Reset Autosampler
The autosampler car has failed to initialize properly. A reset is required.
Corrective Action
1
Open the autosampler access door.
2
Close the autosampler access door. The autosampler will reset.
3
If the problem persists, press the Off button on the analyzer touchpad.
4
Wait about 1 minute and press the On button to restart the analyzer.
5
If the problem persists, call Siemens Service for assistance.
Autosampler - Car Lost Steps, Reset Autosampler
The autosampler car is not in the expected position. A reset is required.
7-16
CAUTION
Corrective Action
Using Exerciser tab, calibrate
Mixer-Shuttle
Using Exerciser tab, calibrate
Mixer-Shuttle
Car Index
position, and
Mixer-Aspirate
Car Index
position, and
Mixer-Aspirate
position,
position.
position,
position.