Siemens ADVIA 2120 Operator's Manual page 197

Table of Contents

Advertisement

connected or is
faulty.
Autosampler - Mode of Operation Not Set
The autosampler was unable to carry out a requested motion while the system
was in the Exerciser tab.
Corrective Action
In the
you want to perform. You may have to reset the autosampler before proceeding.
To reset the autosampler
1
At the analyzer touchpad, select
2
Wait about 1 minute and select
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Motion Denied, Reset Autosampler
The autosampler has failed to execute a command to move and must be reset.
Corrective Action
1
At the analyzer touchpad, select
2
Wait about 1 minute and select
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Motion Request Failed, Reset Autosampler
The autosampler failed to perform a requested motion and might be defective. An
autosampler reset is required.
Corrective Action
1
Select
2
Wait about 1 minute and select
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Motor Controller Failed, Reset Autosampler
The autosampler motor controller failed to respond to a routine command and
might be defective. An autosampler reset is required.
Corrective Action
1
Select Off on the analyzer touchpad.
7-28
tab, set the autosampler to the correct mode for the operation
Exerciser
at the analyzer touchpad.
Off
Off
.
to restart the analyzer.
On
Off
.
to restart the analyzer.
On
On
to restart the analyzer.