NOTE
Removing the slide rack loader will cause an
message will clear when you perform an Autoslide Mechanical Initialization
at the end of this procedure.
3. Holding both sides of the printer cartridge, press toward the inside of the
Autoslide and pull the cartridge out of its holder. Be careful not to tangle the
ribbon when removing it.
4. Install the new printer cartridge securely in its holder. Make sure the notches
on the cartridge are positioned correctly in the holder.
5. Replace the slide rack loader, close the lower front cover.
6. Place the analyzer in Ready mode, then turn the Autoslide on. Wait a few
seconds. The Autoslide will perform a Mechanical Initialization. You must
wait for the initialization to execute automatically.

Replacing the Autoslide Fuses

IMPORTANT
Do not attempt to replace the Autoslide fuses.
For replacement of all Autoslide fuses, call your Siemens technical support
provider or distributor.
10-46
condition. This
ASL Stop