Siemens ARCADIS Varic Start-Up
Hide thumbs Also See for ARCADIS Varic:

Advertisement

Quick Links

M ü h l h ä u s e r
ARCADIS Varic
Start-up
System
Start-up Instructions
Print No.:
SPR2-310.815.01.05.02
Replaces: SPR2-310.815.01.03.02
08080017
SP
2004
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 02.05
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens ARCADIS Varic

  • Page 1 M ü h l h ä u s e r ARCADIS Varic Start-up System Start-up Instructions 08080017 2004 © Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages.
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Sony printer LPR-1000MD ..........37 Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 3 of 48 Medical Solutions 02.05...
  • Page 4 Completing protocols..........39 Completing the "ARCADIS Varic Installation" protocol ..... . . 39 5 _______ Protocols _____________________________________________________ 40 C-arm movement and emergency stop.
  • Page 5: Additionally Required Documents

    General information Additionally required documents 1General information • ARCADIS Varic operating instructions • ARCADIS Varic system wiring diagram • IQ test (quality assurance) • Image intensifier laser light localizer (optional) assembly and setting instructions Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 5 of 48 Medical Solutions 02.05...
  • Page 6: Tools, Aids (Partially For Iq Test)

    51 38 727 Y0766 e.g. Unimed 1100 (Leakage current measurement only required in the case of a deviation of the local line voltage from the system delivery sta- tus). ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 6 of 48 02.05 CS SD 24...
  • Page 7: Note On The Log Book

    The log book is located in the monitor cart behind the keyboard. It is accessible if the cover on the back of the monitor cart is removed. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 7 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 8: Notes On Adapting The Power Plug To Local Conditions

    Varic (refer to current and voltage values stated on the line voltage label, back of moni- tor cart). • The required line internal resistance is reached (refer to ARCADIS Varic Planning Guide). • The selectivity of the fuse protection is assured - according to the relevant national standards.
  • Page 9: Note On The Power Connection

    The system delivery state is provided by the power connection values on the the line volt- age label (back of the monitor cart). Necessary adjustments to local conditions are to be performed as necessary according to the adjustment instructions of the ARCADIS Varic system.
  • Page 10: Notes On Equivalent Leakage Current Measurement

    VDE 0751, Part 1 (see also ARTD, Part 2). In addition, they may not exceed the initial measured value by more than 50%. The system must be repaired if the limit is exceeded. Document the measured value in the equivalent leakage current protocol. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 10 of 48 02.05...
  • Page 11: Safety Instructions

    General information Safety instructions • When performing the work steps and checks, the general safety information for medi- cal products must be observed. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 11 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 12: Switching The System To Be Free Of Voltage

    Switching the system to be free of voltage • To switch the system to be free of voltage in all its parts, bring the system switch into the OFF position. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 12 of 48 02.05...
  • Page 13: Parts Conducting Voltage After System Off And/Or With Power Plug Pulled Out

    Therefore before working on one of these boards, make sure that the LED V400 has gone out. This is the case 3 minutes after the ARCADIS Varic system is switched off. Monitor cart, monitors and image system WARNING...
  • Page 14: Power Connection

    Power connection 2General Start-up Measuring the line voltage • Measure the voltage and frequency at the intended outlet for the ARCADIS Varic sys- tem. • Compare the measured line voltage with the line voltage label on the back of the moni- tor cart.
  • Page 15: The Line Voltage Can Be Adapted At The Transformer In The Monitor Cart

    Pull the power plug out of the UPS in the monitor cart. • Adapt the voltage by changing the connections to transformers T1 and T2. Refer to the ARCADIS Varic wiring diagram. Fig. 1: Monitor cart line voltage Pos. 1 Power plug Pos.
  • Page 16: Measuring The System Leakage Current

    General Start-up Measuring the system leakage current NOTE Perform the system leakage current measurement only if the line voltage on the ARCADIS Varic system must be adapted to the local line voltage. also (Notes equivalent leakage current measurement / p. 10)
  • Page 17: 1990 (E.g. Bender Tester)

    • Measurement configuration according to (Fig. 2 / p. 16) application part (1/Fig. 3 / p. 17) is not available in ARCADIS Varic. • For safety reasons place the system on an insulated base. • Perform measurement and repeat measurements.
  • Page 18: Function Check Of The Emergency Stop Switches

    "The emergency stop switch has been released." • Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm movement and emergency stop (C-arm movement and emergency stop / p. 40). ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 18 of 48 02.05 CS SD 24...
  • Page 19: Checking The Temperature Indicator

    If the inner, square field of the indicator is white, the temperature has not been exceeded. Remove the temperature indicator. If the indicator is discolored (inner field black), proceed according to IQ docu- ment RXD0-000-038.01.xx.xx. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 19 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 20: Function Checks

    General Start-up Function checks Operation of the ARCADIS Varic, see operating instructions • Stand and C-arm system movements • Semi-transparent slot and iris diaphragm in fluoroscopy procedure. • Collimation of the semi-transparent slot and iris diaphragm using fluoroscopy proce- dure.
  • Page 21: Testing The Ground Wire

    10 A) • Test procedure: Measure between all touchable, conductive parts of the system and the ground wire at the ARCADIS Orbic power plug. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 21 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 22: Functional Test Of C-Arm Lift Movement And Parking Brakes

    Document the function of the C-arm movements, brakes and warning signals in the test protocol, "C-arm movement and emergency stop" (C-arm movement and emergency stop / p. 40). ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 22 of 48 02.05 CS SD 24...
  • Page 23: System Configuration

    Please clarify previously with the customer whether the following system configura- tions are required. Service login: Fig. 5: Service login • Click on <Options> <Customer service> <Local service>. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 23 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 24 Specific system settings Fig. 6: Authentication_ • Enter the 6-character password and click OK (see system folder for password). ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 24 of 48 02.05 CS SD 24 Medical Solutions...
  • Page 25: Customer Address Configuration

    Specific system settings Customer address configuration Fig. 7: Configuration_ • Click on <Configuration>. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 25 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 26 Specific system settings Fig. 8: Configuration • Click <Next>. Fig. 9: Configuration_Local Host • Click <Site Info>. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 26 of 48 02.05 CS SD 24 Medical Solutions...
  • Page 27: Main System Configuration

    "serial no." or "station name". • Click <Save>. • Click <Home>. Main system configuration • Start the service program and log in as described under "Service Login". Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 27 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 28 Specific system settings Fig. 11: Main_system • Click <Main system>. Fig. 12: Main system_ • Select ARCADIS Varic and click <Next>. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 28 of 48 02.05 CS SD 24 Medical Solutions...
  • Page 29 - Buzzer on: Buzzer sound always during radiation. The country-specific regulations must be observed for the max. fluoroscopy time and the radiation blockage. Factory setting: Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 29 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 30: Country-Specific Adaptation Of The Maximum Tube Current

    NOTE A maximum tube current limit is required - as far as known - only in Denmark. • Start service program and log in as described under "Service Login". ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 30 of 48 02.05 CS SD 24...
  • Page 31 Specific system settings Fig. 14: Main_system • Click <Main system>. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 31 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 32 Specific system settings Fig. 15: Main system • Select ARCADIS Varic and click <Next>. Fig. 16: Main System_Configuration_Current Limitation • Click <Current limitation>. • Make country-specific changes (only Denmark) and click <Save>. • Click <Home>. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 32 of 48 02.05...
  • Page 33: Backup

    Select "Backup & Restore" in the “Service Software Home Menu” • Place the backup CD (located in the monitor cart service compartment) in the CD ROM drive. Fig. 17: Backup & Restore_Command Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 33 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 34 Select the <Backup> command. Fig. 18: Backup & Restore_Command_Backup • Select the respective drive under “Drives”. - Save to CD ROM (standard save) Fig. 19: Backup & Restore_Command_Backup_Packages ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 34 of 48 02.05 CS SD 24...
  • Page 35 - Wait until the message “Ready” is displayed in the footer MainSystem • Select “MainSystem” under Packages. • Click <Go> - Wait until the message “Ready” is displayed in the footer Exiting the backup Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 35 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 36 Store the CD ROM with the other system documentation. NOTE It is not possible to back up the database (patient images)! Store the patient images in an archive. Saving to CD ROM as a long-term archive is prohibited. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 36 of 48 02.05...
  • Page 37: Paper Printer (Optional)

    Perform the following steps according to the user manual for the printer (CD-ROM): • Unpack the printer. • Fill the paper tray with paper (follow instructions on paper package). • Plug in the USB cable. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 37 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 38 Final Work Steps • Print a test page after the automatic test. • Follow the instructions on the CD-ROM from the manufacturer in the event of problems. ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 38 of 48 02.05 CS SD 24...
  • Page 39: Final Work Steps

    Configuring the network (optional) and remote service Configure the network according to the Configuration Guide, SPR2-310.843.01.01.xx.xx and the remote connection according to Installation, Siemens Remote Service, SP00-000.816.02.01.xx.xx. Completing protocols After completing all adjustment work steps and check measurements, the responsible technician must sign and date the test protocols and country-specific protocols to confirm that all values have been correctly determined and recorded.
  • Page 40: C-Arm Movement And Emergency Stop

    Protocols C-arm movement and emergency stop 5Protocols System: ARCADIS Varic Material no.: ................. Serial number: ................. Tab. 1 C-arm movement and emergency stop Movement Name Date Initials Yes/No Releasing/locking the brakes Downward movement, position Downward movement, position The warning signals sound as described.
  • Page 41: Leakage Current Measurement

    Protocols Leakage current measurement System: ARCADIS Varic Material no.: ................. Serial number: ................. Measuring arrangement see reverse: Tab. 2 Leakage current measurement Equivalent Measuring Name Date Initials instrument Unit leakage used, serial current Initial measured value Repeat measurement Repeat measurement...
  • Page 42 Equivalent Measuring Name Date Initials instrument Unit leakage used, serial current Repeat measurement Repeat measurement Repeat measurement Repeat measurement Repeat measurement Repeat measurement Repeat measurement ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 42 of 48 02.05 CS SD 24 Medical Solutions...
  • Page 43: Network Checklist

    Configuration/ DICOM/ General Study transfer * Configuration/ DICOM/ General Print * Configuration/ DICOM/ General Remarks Network data for HIS/RIS network partner (only with HIS/RIS option) Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 43 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 44 Supported DICOM services: Storage Yes * Configuration/ DICOM/ Network nodes Storage commitment Yes * Configuration/ DICOM/ Network nodes Query Node name: * Configuration/ DICOM/ Network nodes ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 44 of 48 02.05 CS SD 24 Medical Solutions...
  • Page 45 * Configuration/ DICOM/ Network nodes Retrieve * Configuration/DICOM/Network nodes Remark: Network checklist (3-) Network data for a DICOM camera in the network (only with the DICOM Print option) Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 45 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 46 General node properties: Logical name * Configuration/ DICOM/ Print devices Application entity: AE title * Configuration/ DICOM/ Print devices Port number * Configuration/DICOM/Print devices Remark: ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 46 of 48 02.05 CS SD 24 Medical Solutions...
  • Page 47 Changes to previous version 6Changes to previous version The revised Completing the "ARCADIS Varic Installation" protocol section was added to the "Final Work Steps" chapter. Siemens AG SPR2-310.815.01.05.02 ARCADIS Varic Page 47 of 48 Medical Solutions 02.05 CS SD 24...
  • Page 48 Changes to previous version ARCADIS Varic SPR2-310.815.01.05.02 Siemens AG Page 48 of 48 02.05 CS SD 24 Medical Solutions...