TROUBLESHOOTING & REPAIR PROCEDURES
11.6.3.2. Metal Wool Scrubber Option
Contact T-API for instructions on replacing the optional Metal Wool Scrubber.
11.6.4. Replacing the IZS O
1. Turn off power to the instrument.
2. Remove instrument cover.
3. The IZS Zero Air Scrubber is attached to the brass elbow inlet fitting on the top
of the O
3
4. Disconnect 1/4" Tube Fitting nut on O3 Generator inlet fitting.
5. Disconnect 1/8" tube fitting on the other end of the scrubber.
6. Install new scrubber by reversing these steps.
268
Generator Assembly. See Figure 11-6.
Figure 11-6: IZS Zero Air Sc rubber Location
Scrubber
3
IZS Zero Air Scrubber
Model 400E Ozone Analyzer
Instruction Manual