Figure 10-6: Photometer Uv Lamp Power Supply Block Diagram - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

THEORY OF OPERATION
UV Lamp Power Supply
+15 VDC
2
I
C Bus
Figure 10-6: Photometer UV Lamp Powe r Supply Block Diagram
UV Lamp Temperature Control
A thermistor and DC heater attached to the UV Lamp to maintain the Lamp at an
optimum operating temperature.
Sample Gas Temperature Sensor
A thermistor attached to the quartz tube for measuring sample gas temperature.
Photometer UV Detector
The vacuum diode, UV detector that converts UV light to a DC current.
UV Detector Preamplifier
A preamplifier assembly, which convert the Detector's current output into a DC
Voltage than amplifies it to a level readable by the Auto D Converter circuitry of the
instrument's mother board.
230
Transformer
Controller
DC
Regulator
Transformer
Error
Amplifier
D-to-A
Converter
Model 400E Ozone Analyzer
Center
Tapped
Rectifier
Instruction Manual
UV Lamp
.
P/N 04316 Rev: B