Calibration The Izs Option O; Generator; Table 6-13: Current Loop Output Check - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Operating Instructions
An alternative method for setting up the Current Loop outputs is to connect a
250 ohm ±1% resistor across the current loop output in lieu of the current meter.
Using a Voltmeter connected across the resistor follow the procedure above but
adjust the output for the following values:
% FS
0
100

6.7.4. Calibration the IZS Option O

This function sets the IZS O
and full scale, measures the actual O
generator lamp drive voltage and generator's O
Whenever a certain O
data in this table to interpolate the correct drive voltage for the desired O
Because the instrument waits 5–7 minutes at each step for
the O
108

Table 6-13: Current Loop Output Check

Voltage across Resistor for
2-20 mA
0.5 VDC
5.0
Generator output to a series of levels between zero
3
output level is requested, the instrument's CPU uses the
3
level to stabilize, this calibration operation often takes
3
more than one hour to complete.
Voltage across Resistor for

Generator

3
output at each level then records the
3
output level in a lookup table.
3
NOTE
Model 400E Ozone Analyzer
Instruction Manual
4-20 mA
1 VDC
5.0
3
P/N 04316 Rev: B
output.