MAINTENANCE SCHEDULE & PROCEDURES

9.3.6. Flow Calibration

Once an accurate measurement has been recorded by the method described above,
adjust the analyzer's internal flow sensors by pressing:
Exit at Any
Time to Return
to main the
SETUP Menu
Toggle these keys until
the displayed flow rate
equals the flow rate being
measured by the
independent flow meter.
214
SAMPLE
RANGE = 500.0 PPB
< TST TST > CAL
SETUP X.X
CFG ACAL DAS RNGE PASS CLK
SETUP X.X
DIAG
COMM VARS
HALT
SETUP X.X
ENTER DIAG PASS: 818
8
1
8
DIAG
SIGNAL I / O
NEXT
Repeat Pressing NEXT until . .
FLOW CALIBRATION
DIAG
PREV NEXT
DIAG FCAL
ACTUAL FLOW: 780 CC / M
0
7
8
0
Model 400E Ozone Analyzer
O3 =XXX.X
SETUP
MORE
EXIT
EXIT
ENTR
EXIT
ENTR
EXIT
.
ENTR EXIT
ENTR EXIT
Instruction Manual
Exit returns
to the
previous menu
ENTR accepts the
new value and
returns to the
previous menu
E XIT ignores the
new value and
returns to the
previous menu
P/N 04316 Rev: B