Pneumatic Connections; Figure 3-2: Rear Panel Connectors - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual

3.1.2. Pneumatic Connections:

1. Sample and calibration gases should only come into contact with PTFE (Teflon),
FEP, glass, stainless steel or brass.
Cooling
Fan
AC Power
Receptacle
P/N 04316 Rev: B
OZONE (O3) IS A TOXIC GAS.
Obtain a Material Safety Data Sheet (MSDS) for this material.
Read and rigorously follow the safety guidelines described
there.
Do not vent calibration gas and sample gas into enclosed areas.
Serial I/O
LED's
Sample
Inlet
Exhaust
Outlet
Span Gas
Inlet
Zero Air
Inlet
Dry Air
Inlet
DCE – DTE
COM A Connector
Switch

Figure 3-2: Rear Panel Connectors

In order to prevent dust from getting into the gas flow channels
of your analyzer, it was shipped with small plugs inserted into
each of the pneumatic fittings on the back panel.
Make sure that all of these dust plugs are removed before
attaching exhaust and supply gas lines.
CAUTION
Status
Outputs
COM B Connector
(RS232 Only)
(RS - 232 or RS - 485)
CAUTION
Getting Started
Analog
Outputs
Control
Inputs
29