Table 10-1: Relay Board Status Led's - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

THEORY OF OPERATION
Status LED's
Eight LED's are located on the Analyzer's Relay board to show the current status on
the various control functions performed by the Relay Board. They are:
LED
Color
D1
RED
1
D2
YELLOW
D3
YELLOW
D4
YELLOW
D5
YELLOW
D6
YELLOW
D7
GREEN
D8
GREEN
D9
GREEN
D10
GREEN
D11
GREEN
D12
GREEN
D13
GREEN
D14
GREEN
D15
GREEN
D16
GREEN
Watch Dog Circuitry
Special circuitry on the Relay Board watches the status of LED D1. Should this LED
ever stay ON or OFF for 30 seconds, the Watchdog Circuit will automatically shut of
all valves as well as turn off the UV Source(s) and all heaters. The Sample Pump
will still be running.
232

Table 10-1: Relay Board Status LED's

Function
Watchdog Circuit
Cycles On/Off Every 3 Seconds under direct control of the
analyzer's CPU.
HEATING
Metal Wool
Scrubber Heater
Spare
N/A
Spare
N/A
Spare
N/A
Spare
N/A
Zero/Span Gas
Valve Open to SPAN GAS
Valve
FLOW
Measure/Ref
Valve Open to REFERENCE
Valve
gas path
Sample/Cal Gas
Valve Open to CAL GAS
Valve
FLOW
Spare
N/A
Spare
N/A
Spare
N/A
Spare
N/A
Spare
N/A
HEATING
Photometer UV
Lamp Heater
HEATING
IZS O
Generator
3
UV Lamp Heater
Status When Lit
Model 400E Ozone Analyzer
Instruction Manual
Status When Unlit
NOT HEATING
N/A
N/A
N/A
N/A
Valve Open to ZERO GAS
FLOW
Valve Open to MEASURE gas
path
Valve Open to SAMPLE GAS
FLOW
N/A
N/A
N/A
N/A
N/A
NOT HEATING
NOT HEATING
P/N 04316 Rev: B

Advertisement

Table of Contents
loading

Table of Contents