6.9.9. Manually Configuring the Network IP Addresses

There are several circumstances when you may need to manually configure the
interface settings of the analyzer's Ethernet card. The INET sub-menu may also be
used to edit the Ethernet card's configuration properties
Your LAN is not running a DHCP software package,
The DHCP software is unable to initialize the analyzer's interface;
You wish to program the interface with a specific set of IP addresses that
may not be the ones automatically chosen by DHCP.
Editing the Ethernet Interface properties is a two step process.
STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to manually set
INSTRUMENT IP, GATEWAY IP and SUBNET MASK is disabled
SAMPLE
RANGE = 50.0 PPM
< TST TST > CAL
SAMPLE
ENTER SETUP PASS : 818
8
1
8
PRIMARY SETUP MENU
SETUP X.X
CFG DAS RNGE PASS CLK
SECONDARY SETUP MENU
SETUP X.X
COMM
VARS DIAG
130
CO =XX.X
SETUP
ENTR
EXIT
MORE
EXIT
EXIT
Continue with editing of Ethernet interface
properties (see Step 2, below).
Model 400E Ozone Analyzer
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
SETUP X.X
DHCP: ON
<SET SET>
EDIT
SETUP X.X
DHCP: ON
ON
SETUP X.X
DHCP: ON
OFF
Instruction Manual
EXIT
EXIT
ENTR EXIT
ENTR EXIT
ENTR accept
new settings
EXIT ignores
new settings