Table 6-12: Voltage Tolerances For Analog Output Calibration - Teledyne 400E Instruction Manual

Ozone analyzer
Table of Contents

Advertisement

Model 400E Ozone Analyzer
Instruction Manual
6.6.2.2. Manual Analog Output Calibration and Voltage
For highest accuracy, the voltages of the analog outputs can be manually
calibrated. Calibration is done through the instrument software with a voltmeter
connected across the output terminals (Figure 6-5). Adjustments are made using
the front panel keys by setting the zero-point first and then the span-point (Table
6-12).
To select manual output calibration for a particular channel, activate the ANALOG
I/O CONFIGURATION MENU, then press:
DIAG
Exit to return to
the main
PREV
sample display
DIAG AIO
< SET
Press SET> to select the analog output channel to
DIAG AIO
< SET SET>
DIAG AIO
Now the analog output channels should either be automatically calibrated or they
should be set to manual calibration, which is described next.
The software allows this adjustment to be made in 100, 10 or 1 count increments.

Table 6-12: Voltage Tolerances for Analog Output Calibration

FULL
SCALE
0.1 VDC
1 VDC
5 VDC
10 VDC
P/N 04316 Rev: B
Adjustment
ANALOG I / O CONFIGURATION
NEXT
ENTR
AOUTS CALIBRATED: NO
SET>
CAL
be configured. Then press EDIT to continue
CONC_OUT_2:5V, CAL
EDIT
CONC_OUT_2 RANGE: 5V
SET>
EDIT
ZERO
TOLERANCE
±0.0005V
±0.001V
±0.002V
±0.004V
DIAG AIO
EXIT
SET>
< SET
DIAG AIO
EXIT
< SET
SET> EDIT
DIAG AIO
ON
EXIT
Toggles the
auto cal mode
ON/ OFF for
this analog
output channel
only.
EXIT
SPAN
SPAN TOLERANCE
VOLTAGE
90 mV
900 mV
4500 mV
4500 mV
Operating Instructions
CONC_OUT_2 REC OFS: 0 mV
EDIT
CONC_OUT_2 AUTO CAL: ON
AOUT AUTO CAL: ON
ENTR EXIT
ENTR accepts the new setting
and returns to the previous menu.
EXIT ignores the new setting and
returns to the previous menu.
±0.001V
±0.001V
±0.003V
±0.006V
EXIT
EXIT
87