Default Values - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

NOTE
When setting calibration timing requirements, bear the following precautions in mind:
(1) If the calibration interval is shorter than the sum of Hold (output stabilization) time plus
calibration time, the second calibration start time will conflict with the first calibration. In such
a case, the second calibration will not be conducted. (When both zero and span calibrations
are to be performed, the calibration time is double that required for a single (zero or span)
calibration.)
(2) For the same reason, if the calibration start time conflicts with manual calibration or semi-
automatic calibration, the current calibration will not be conducted.
(3) If the calibration time conflicts with maintenance service or blow back operations, calibration
will start after completing the maintenance service or blow back operations (see Section
8.2.1, earlier in this manual).
(4) If 000 days, 00 hours are set for the calibration intervals, only the first calibration will be
conducted; a second or later calibration will not be conducted.
(5) If a past date is set to the calibration start day, no calibration will be conducted.
9.2.6

Default Values

When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as
shown in Table 9.1.
Table 9.1
Calibration mode
Automatic calibration procedure
Zero gas concentration
Span gas concentration
Hold time [mm:ss]
Calibration time [mm:ss]
Interval [day]
Interval [hour]
First start date [YY/MM/DD]
First start time [hh:mm]
Default Settings for Calibration
Item
Manual
Span and Zero
1.00%O
21.00%O
10:00
10:00
30
0
00/01/01
00:00
Default Setting
2
2
<9. Calibration>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
9-9

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr802gZr202g

Table of Contents