Response Time: Response of 90% within 5 seconds. (Measured after gas is introduced from
calibration gas inlet and analog output starts changing.)
Safety, EMC, and RoHS conformity standards of ZR22G, ZR802G
Installation altitude: 2000 m or less
Installation category: (IEC61010); II
Pollution degree: (IEC61010); 2
Note
· Installation category, called overvoltage category, specifies impulse withstanding voltage. Category II is for
electrical equipment.
· Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which reduce dielectric
strength. Degree 2 is the normal indoor environment.
Safety:
ZR22G;
CE
EN 61010-1
EN 61010-2-030
UL
UL61010-1:2nd Edition
CSA
CAN/CSA-C22.2 No. 61010-1
GB
GB30439 Part 1
ZR802G;
CE
EN 61010-1/A1
EN 61010-2-030
UL
UL61010-1:3rd Edition, AMD1
UL61010-2-030:1st Edition
CSA
CAN/CSA-C22.2 No. 61010-1+Amd1
CAN/CSA-C22.2 No. 61010-2-030
GB
GB30439 Part 1
EMC:
CE
EN 61326-1 Class A Table 2
EN 61326-2-3,
EN 61000-3-2
RCM
EN 55011 Class A, Group 1
KC
KN11 Class A Group1, KN61000-6-2
Note
· This instrument is a Class A product, and it is designed for use in the industrial environment. Please use this
instrument in the industrial environment only.
· Influence of immunity environment (Criteria A) : Output shift is specified within ±20% of F.S.
RoHS:
ZR22G; EN 50581
ZR802G; EN IEC 63000
Others:
REACH Regulation EC 1907/2006
Information of the WEEE Directive
only and, therefore, is out of scope of the WEEE Directive. The WEEE Directive is only
valid in the EU.
This product is purposely designed to be used in a large scale fixed installations
<2. Specifications>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
2-3