8.7.6

Setting Purging

Purging is to remove condensed water in the calibration gas pipe by supplying a span calibration
gas for a given length of time before warm-up of the detector. This prevents cell breakage during
calibration due to condensed water in the pipe.
Open the solenoid valve for the automatic calibration span gas during purging and after the purge
time has elapsed, close the valve to start warm-up.
Purging is enabled when the cell temperature is 100°C or below upon power up and the purge
time is set in the range of 1 to 60 minutes.
How to setup purging
(1)
"Converter menu" > "Setting"
(2)
Select "Others setting". Enter "Purging time" value.
Figure 8.19
8.7.7

Setting password

The instrument can protect various setup and executes by means of a password. You can
setup the password of "Commissioning" and "Execute" respectively for each item as below.
"Commissioning" password is for setting each revised setup data. "Execute" password is to
execute maintenance or calibration.
Table 8.19
Menu
Converter menu Detail
Sensor menu
NOTE
You can establish password for the calibration or a shortcut of Favorite on Home screen as well.
Purging time setting
Pasword Setting Items
Item
(Commissioning / Execute)
-
Trend
-
Blowback
Execute
Others
Commissioning
Maintenance
Execute
Setting
Commissioning
Reset
Commissioning
Lang.
Execute
Detail
-
Calibration
Execute
Cell resist
Execute
Setting
Commissioning
<8. Detailed Data Setting>
Password
IM 11M12G01-02EN
8-28
1sh Edition : Mar. 25, 2021-00

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr802gZr202g

Table of Contents