Output Range Setting - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

Figure 7.5
7.8

Output Range Setting

This section sets forth analog output range settings. For details, consult Section "8.1
Output Setting", later in this manual.
Minimum Current (4 mA) and Maximum Current (20 mA) Settings
To set the minimum and maximum current settings, follow these steps:
(1) Select the "Setting" from Converter menu.
(2) Select "mA-output settings".
(3) Select "mA-output1".
(4) On "Selection of AO1 Oxygen concentration" enter "4mA point" and "20 mA point"
(5) Set "mA-output2" in the same manner as steps above.
Figure 7.6
NOTE
Each setting is limited in value. See "8.1 Current Output Setting" for details.
Choise of moisture base
Setting "mA-output"
<7. Startup>
Current
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
7-5