Save Load; Log File Output - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

(1) "Converter menu" > "Maintenance" > "Communication settings."
(2) Select "TCP settings" to setup.
Figure 10.24
Note
Setup of Ethernet is reflected after ZR802G is restarted.
10.8

Save load

Converter has functions for outputting log files, setup load, and updating software using SD-Card.
Precautions When Using SD Card
l
• Be sure to format the SD card with SD Association software. If you use an SD card without
being formatted by the tool, operation is not guaranteed.
To download the formatter, select [Download] from the link below and proceed to SD
Memory Card Formatter.
https://www.sdcard.org
• Do not disconnect Power Supply or remove the SD card while accessing (reading/writing,
software updating) the SD card.
10.8.1

Log file output

You can output the following files as log files:
• Maintenance report
Outputs the last three calibration and various set values.
• Measured value log
The following log data is output. Log cycle and measurement days can be selected from "1
second cycle × 8 days", "2 second cycle × 16 days", and "5 second cycle × 40 day
• Date and time
• Oxygen concentration
• Cell voltage
• CJ temperature
• Thermocouple temperature
• Simple cell resistance value
• NE107 Status
• Equipment status
TCP settings
<10. Other Functions>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
10-18