YOKOGAWA ZR22G User Manual page 150

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

How to setup the measured gas pressure
(8) Enter the pressure (absolute pressure) of the measured gas in "Exhaust gas pressure."
Figure 8.18
Setting Measurement Gas Temperature and Pressure (Humidity Analyzer)
NOTE
• The critical temperature of the saturated water vapor pressure is 374°C. If a gas
temperature exceeding 370°C is entered, no correct calculation will be obtained.
• If an invalid value is set, no correct calculation will be obtained. Be sure to check allowable
temperature ranges of the temperature transmitter you use, and then enter the value
correctly.
Default Values
l
When the analyzer is delivered, or if data are initialized, measurement gas temperature and
pressure set ranges and their default settings are as shown in Table 8.18.
Table 8.18
Measurement Gas Temperature and Pressure Set Ranges and Default Settings
Set Item
Temperature input selection
Input temperature setting
4 mA input temperature value
20 mA input temperature value
Temp. upper limit alarm value
Temp. lower limit alarm value
Exhaust gas pressure
<8. Detailed Data Setting>
Set range
Default setting
-
Preset value
0 to 3000°C
0 to 3000°C
0 to 3000°C
0 to 3000°C
0 to 3000°C
0 to 689.47 kPa abs
101.33 kPa abs
300°C
0°C
1000°C
1000°C
0°C
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
8-27