10.5.3

Setting Output Hold Time and Blow back Time

If the blow back mode is in "No function", the output "Hold time" and "Blow back time" are not
displayed. If you select "Hold time", the numeric-data entry display appears. Enter the desired
"Hold time" (output stabilization time) from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
When you select "Blow back time", the numeric-data entry display appears. Enter the desired
"Blow back time" from 00 minutes, 00 seconds to 60 minutes, 59 seconds.
10.5.4

Setting Interval, Start Date, and Start Time

The "Interval" is the time to execute blow back automatically. Display the numeric-data entry
panel display to set the desired interval from 000 days, 00 hours to 255 days, 23 hours.
For the "Start date" and "Start time", set the date when the blow back is first executed and the
time when to start the blow back, respectively. If you want to execute the first blow back, for
example, at 4:00 p.m. on March 25, 2001, enter 25/ 03/ 01 for the Start date and 16:00 for the
Start time.
Figure 10.19
In the Blow back setup display shown in Figure 10.10, if you choose "Mode: No function" or
"Semi_Auto", the Interval, Start date, and Start time for these are not displayed.
NOTE
• If the blow back is executed with an input contact, it must be preset in the Input contact
settings (for more details, see Section "8.6 Input Contact Settings").
• Set the Output contact used as the blowback switch (for more details, see Section "8.5
Output Contact Setup").
• Do not set any other function for the contact used as the blow back switch.
Otherwise, blow back may be activated when the contact is closed by any other function.
• No blow back is executed during calibration. If the start time of the automatic blow back
comes, the calibration ends and the blow back will be executed after the instrument returns
to the measurement status.
• If the start time of the automatic blow back comes during a maintenance mode or semi-auto
blow back, the auto blow back of this time will not be executed.
• If you set the blow back interval at 000 days, 00 hours, only the first blow back is then
executed. No subsequent blow backs will be executed.
• If a past date is set for the Start time, no automatic blow back will be executed.
Setting Interval, Start Date, and Start Time
<10. Other Functions>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
10-13