YOKOGAWA ZR22G User Manual page 148

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

(4) When you select "Pressure input selection", a dialog box appears to select "Preset value",
"External input". Select according to your system.
When Preset value is selected
(5) When "Preset value" is selected on "Pressure input selection", enter a sample gas pressure
on "Input pressure set value".
When External input is selected
(6) When "External input" is selected on "Pressure input selection", Enter the values of pressure
of 4 mA point and 20 mA point respectively of a pressure transmitter you use.
(7) When alarm from sample gas pressure is used, select "Pressure upper limit alarm value",
"Pressure lower limit alarm value." Enter the alarm pressure value with numeric input.
Figure 8.17
Setting Measurement Gas Pressure (Oxygen Analyzer)
Default Values
l
When the analyzer is delivered, or if data are initialized, measurement gas temperature and
pressure set ranges and their default settings are as shown in Table 8.17.
Table 8.17
Measurement Gas Pressure Set Ranges and Default Settings
Set Item
Pressure input selection
Input pressure set value
4 mA input pressure value
20 mA input pressure value
Pressure upper limit alarm value
Pressure lower limit alarm value
<8. Detailed Data Setting>
Set range
-
-500 to 500 kPaG
-500 to 500 kPaG
-500 to 500 kPaG
-500 to 500 kPaG
-500 to 500 kPaG
Default setting
Preset value
0.00 kPaG
-5.00 kPaG
5.00 kPaG
5.00 kPaG
-5.00 kPaG
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
8-25