Setup Of Hold Time, Interval, Start Date, And Start Time; Default - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

10.6.2

Setup of Hold time, interval, start date, and start time

The stabilization time is turned setup because the measured value changes temporarily
immediately after the simple cell resistance measurement. When "Hold time" is selected, the
numeric entry screen is turned Display, so enter the hold time. Setup can range from 00 minutes
00 seconds to 60 minutes 59 seconds. Default has been setup in one minute. If no significant cell
degradation occurs, the meter returns to the normal measurement in 1 minute.
When using Auto Simple Cell Resistance Measure, "Interval", "First Start Date" and "First Start
time" must be turned setup. The term "Interval" refers to the period at which automatic simple
cell resistance measurement is performed. Select "Interval" and enter it from the numeric entry
screen. Setup is possible for 000 days 00 hours to 255 days 23 hours. For Start Date and Start
Time, setup the date on which you want to perform the first simple cell resistance measurement
and the start time.
Note
• The degradation tendency of the cell can be checked with the simple cell resistance trend
by performing automatic simple cell resistance measurement. Setup of the automatic
simple cell resistance measurement is recommended when the measurement downtime by
executing the simple cell resistance measurement is not an issue.
• A measurement time of approximately 15 seconds is usually required for simple cell
resistance measurement. If the hold time is included, the measurement stops for more than
1 minute.
• Simple cell resistance measurement is not executed during calibration. When the start time
of auto-simple resistance measurement comes during calibration, the calibration ends and
the simple resistance measurement will be executed after the instrument returns to the
measurement status.
• When the start time of auto-simple cell resistance measurement comes during
maintenance, semi-auto simple cell resistance measurement, the auto simple cell
resistance measurement of this time will not be executed.
• When "000 days 00 hours" is set to setup for the interval, simple cell resistance
measurement is executed only for the first time and is not performed after the second time.
• If the previous date is turned setup on the starting date, the auto simple cell resistance
measurement is not executed.
10.6.3

Default

When the analyzer is delivered, or if data are initialized, the default set values are as shown in
Table 10.7.
Figure 10.7
Item
Mode
Hold time [mm:ss]
Interval [day]
Interval [hour]
First start date [YY/MM/DD]
First start time [hh:mm]
Simple cell resistance alarm value
Default value of the simple cell resistance setting
Semi-automatic
01:00
1
0
00/01/01
00:00
2000 Ω
<10. Other Functions>
Default
IM 11M12G01-02EN
10-15
1sh Edition : Mar. 25, 2021-00