10.4.2

NE107 mode

You can change the alarm display according to the NAMUR NE 107.
(1) "Converter" > "Maintenance" > "Display setting" > "Others setting"
(2) Select NE107 Mode. Select ON/Off.
Table 10.4
Icon
Failure (no power supply to heater)
Function Check, Out of Specification, Maintenance Required
Table 10.5
Icon
Failure (no power supply to heater)
Function Check
Out of Specification
Maintenance Required
10.4.3

Backlight time

You can set the back light to turn off automatically to keep LED life long. The backlight will turn off
if no screen action is taken during a setup time.
(1)
Select "Maintenance" → "Display setup" → "Other setup" from "converter Menu" screen.
(2)
Select "Backlight time". You can setup "disable", "10 minutes", "30 minutes" or "60
minutes"
10.4.4

Entering Tag Names

You can assign any tag name to the instrument;
If there is an instruction in the specification at the time of order, it is entered at the time of
shipment.
(1) Select "Maintenance" → "Display settings" → "Display Items" from "converter Menu" screen.
(2) Select "Tag Name" to enter screen and use alphabets, numbers, and symbols. You can
enter up to 32 character of characters.
Other settings Display
NE107 - Off
Alarm setting
NE107 - ON
Alarm setting