Confirmation of Converter Type Setting

This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has been
set.
CAUTION
If the converter type setting is changed, the operating data that have been set are initialized and
the default settings remain.
(1) Press the [Converter menu] key.
(2) Select [Reset] > [Model setting].
(3) Confirm the displayed product model is one presently being used. When the specification
for an High Temperature Humidity Analyzer is designated at order, High Temperature
Humidity Analyzer is preset at the factory shipment.
(4) To change model, press [Oxygen model] or [Humidity model]. After changing the model,
press [Execute].
(5) If a converter model is changed after setting the operating data, the entered data are
initialized. Enter again the operating data to meet the model to use.
Figure 7.3
Equipment Setup
<7. Startup>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
7-3