8.5
Output Contact Setup
8.5.1
Output Contact
Mechanical relays provide contact outputs. Be sure to observe relay contact ratings. (For details,
see Section "2.1 General Specifications") The operation modes of each contact output are as
follows. For output contacts 1 to 3 you can select open or closed contact when the contact is
"operated." For output contact 4, contact is closed only.
When power fails, contact outputs 1 to 3 are open, and 4 is closed.
Table 8.11
Contact output 1
Contact output 2
Contact output 3
Contact output 4
State when contact "operated"
Open (deenergized) or closed (energized) selectable.
Open (deenergized) or closed (energized) selectable.
Open (deenergized) or closed (energized) selectable.
Closed (deenergized) only
<8. Detailed Data Setting>
When no power is applied
to this equipment
Closed
IM 11M12G01-02EN
8-12
Open
Open
Open
1sh Edition : Mar. 25, 2021-00