In such a case, connect a needle valve (found on the local market) through a nipple (also found
on the local market) to the probe adapter sample gas exhaust (Rc 1/2) so that the sample gas
exhaust volume is restricted.
Reducing nipple
(R1/2-R1/4 or R1/2-1/4NTP)
Needle valve
Figure 4.4
Mounting the needle valve for restricting exhaust flow of the sample gas
In cases where condensation is likely to occur in the probe adapter when the sample gas is
cooled, protect the probe adapter with an insulating material as illustrated in Figure 4.5.
Sample gas outlet
Figure 4.5
Preventing to condensation
If the sample gas is to be vented at a distance from the detector because no reference gas piping
can be provided, an exhaust pipe should be installed as illustrated in Figure 4.6.
In addition, the exhaust pipe shall be kept warm to protect against condensation.
Exhaust pipe
Figure 4.6
Exhaust pipe
F4-4E.ai
Cover flange
Adapter for high temperature probe
Detector
Insulating material
Probe
F4-5E.ai
Rc1/2
Use thick piping for exhaust.
Horizontal part should be kept to the
minimum to maintain piping strength.
F4-6E.ai
<4. Piping>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
4-4