10.8.3

Update software

(1) "Converter menu" > "Other" > "Save/Load"
(2) Select "Update software"
(3) Save a designated file into a SD card to update the software. Normally you don't need to
update the software. If you need the update file, contact Yokogawa.
Figure 10.28
10.9

Data Initialization

Parameter settings can be initialized to the factory default settings. Initialization can be done
for all parameters or for individual parameters. The parameters that can be initialized and their
defaults are listed in Table 10.5.
(1) "Converter menu" > "Initialization"
(2) Select "Factory initialization"
(3) Press "Execute" to initialize to factory default.
Figure 10.29
WARNING
Do not turn off the power during initialization. Otherwise, initialization will not be performed
properly.
Update software
Initialization
<10. Other Functions>
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
10-21