12.2
Displays and Remedies When Alarms are
Generated
12.2.1

Alarm Types

When an alarm is generated, the alarm indication blinks in the display to notify of the alarm.
Pressing the alarm indication displays a description of the alarm.
Alarms include those shown in Table 12.2.
Table 12.2
Alarm
number
101
Oxygen concentration high high alarm
102
Oxygen concentration high alarm
103
Oxygen concentration low alarm
104
Oxygen concentration low low alarm
105
Humidity high high alarm
106
Humidity high alarm
107
Humidity low alarm
108
Humidity low low alarm
109
Mixing ratio high high alarm
110
Mixing ratio high alarm
111
Mixing ratio low alarm
112
Mixing ratio low low alarm
113
Relative humidity high high alarm
114
Relative humidity high alarm
115
Relative humidity low alarm
116
Relative humidity low low alarm
117
Simple cell resistance alarm
118
AO1 saturation
119
AO2 saturation
120
Calibration stability alarm
201
Zero correction ratio high alarm
202
Zero correction ratio low alarm
203
Span correction ratio high alarm
204
Span correction ratio low alarm
205
Cold junction temperature high alarm
206
Cold junction temperature low alarm
207
Thermocouple voltage high alarm
208
Thermocouple voltage low alarm
209
AI current high alarm
210
AI current low alarm
211
Input temperature high alarm
212
Input temperature low alarm
213
Input pressure high alarm
214
Input pressure low alarm
Types of Alarms and Reasons for Occurrence
Alarm Type
<12. Troubleshooting>
Reasons for Occurrence
Oxygen density exceeds setup limit.
Oxygen density exceeds setup limit.
Oxygen level is lower than setup.
Oxygen level is lower than setup.
The water content is above setup limit.
The water content is above setup limit.
The water content is below setup.
The water content is below setup.
Mixing ratio exceeds setup
Mixing ratio exceeds setup
Mixing ratio is lower than setup
Mixing ratio is lower than setup
The relative humidity exceeds setup value.
The relative humidity exceeds setup value.
The relative humidity is less than setup relative humidity.
The relative humidity is less than setup relative humidity.
The simple cell resistor has exceeded setup limit.
mA output has reached the upper or lower limit
mA output has reached the upper or lower limit
Electromotive force of the cell does not stabilize after
calibration period has elapsed.
Zero point correction ratio exceeds 130%
Zero point correction rate is lower than 70%.
Span point correction ratio exceeds 18%
Span point correction rate is lower than-18%.
Temperature of the cold contact has exceeded 155°C.
Temperature of the cold contact is below-25°C.
Thermocouple electromotive force is over 42.1 mV (approx.
1020°C)
Thermocouple electromotive force is below-5 mV (approx.-
170°C)
AI-input mA is above 20.5 mA
AI-input mA is below 3.8 mA
The incoming temperature has exceeded setup threshold.
The incoming temperature is lower than setup.
The input pressure exceeds setup value.
The input pressure is lower than setup value.
IM 11M12G01-02EN