Installation Of The Case Assembly (E7044Kf) - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

4-Φ6.5
Figure 3.19
3.6

Installation of the Case Assembly (E7044KF)

The case assembly is used to store the G7001ZC zero gas cylinders.
The following should be taken into consideration:
(1) Easy access for cylinder replacement
(2) Easy access for checking
(3) Near to the detector or converter as well as the flow setting unit.
(4) The temperature of the case should not exceed 40°C due to rays of the sun or radiated heat.
(5) No vibration
Mounting
n
Mount case assembly on a pipe (nominal JIS 50A) as follows:
(1) Prepare a vertical pipe of sufficient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the case assembly. (The sum of the case assembly and the calibration gas cylinder weighs
approximately 4.2 kg.)
(2) Mount the case assembly on the pipe by tightening the nuts with the U-bolt so that the metal
fitting is firmly attached to the pipe.
Wall Mounting
<3. Installation>
F3-16_1E.ai
IM 11M12G01-02EN
3-12
1sh Edition : Mar. 25, 2021-00