10-35
<10. Other Functions>

10.10 Reboot

Reboot enables the equipment to restart. If the equipment is rebooted, the power is turned off
and then back on. In practical use, the power remains on, and the equipment is restarted under
program control.
When Fault happens, for safety, the power supply to the sensor heater is turned off. To cancel
the error (Fault), reboot by following the procedure below or turn off the power once to restart the
system.
CAUTION
Make sure that before a reboot or restarting the power that there is no problem with the detector
or converter.
After the instrument reboots, if Fault occurs again, turn off the power. Find the problem by trouble
shooting.
How to reboot the system
(1) "Converter menu" > "Others"
(2) "Reboot" A confirmation screen appears. Tap "Execute".
(3) A confirmation screen appears again .Tap "Yes" to reboot.
Figure 10.30
Reboot the system
IM 11M12G01-02EN