YOKOGAWA ZR22G User Manual page 182

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

Selecting Measurement log storage cycle
(1) "Converter menu" > "Setting" > "Other setting"
(2) Select "Measurement log storage cycle" to setup. Default is 2 seconds (16 days).
Figure 10.25
Measurement log storage cycle
• Setup File
You can save a variety of set value. Tag names, passwords, language setup, and
communication addresses are not saved. You can export the data files or copy to other
converter to backup the configuration.
• Event Log
Outputs the data that can be checked in the log data of converter detailed information. This
data is used for checking by our service when a device malfunctions
Table 10.8
File name and output format of log file output
Output file
Maintenance report
Report/
Measured value log
MeasLog/
Setup file (setup save)
SaveLoad/
Event Log
EventLog/
Procedure for outputting log files
(1) "Converter menu" > "Other menu" > "Save/Load."
(2) Select the file that you want to output.
(3) Check that the file name and press "Execute" to export the data into SD card. The SD card
can be connected or disconnected except for when writing or reading files.
Folder name
ZC_report_YYYYMMDDhhmm.csv
ZC_measure_YYYYMMDDhhmm.csv
ZC_setting.*
ZC_event_YYYYMMDDhhmm.L00
<10. Other Functions>
File name
IM 11M12G01-02EN
10-19
Output format
.csv format
.csv format
Binary file format
Binary file format
1sh Edition : Mar. 25, 2021-00

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zr802gZr202g

Table of Contents