5.2 Basic Setup
If the rating of the secondary side of the CT is 5 A, We recommend that
you use a 50 A type clamp-on probe and measure at the 5 A range.
Make sure to set the appropriate VT and CT ratios.
Nominal V
Set the nominal voltage applied from the DUT.
Settings
*
50V~600V(100V
)
* Default value
<Procedure>
Use the up and down arrow keys to move to Nominal V.
Press ENTER
to show a value entry window.*
Use the arrow keys to enter the
nominal voltage.
Press ENTER to confirm or ESC to cancel.
* The input range is shown in a pop-up.

Default Values

When you move to the Nominal V item, you can press F1 (default values)
to show a list of typical nominal voltages that you can choose from.
Settings
100V/101V/110V/120V/200V/202V/208V/220V/230V/
240V/277V/346V/380V/400V/415V/480V/600V
<Procedure>
Use the arrow keys to select nominal voltage.
Press ENTER to
confirm or ESC to cancel.
5-7
IM CW500-01EN