Specifications; General Specifications - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

2.

Specifications

This chapter describes the specifications for the following:
ZR22G
ZO21R
ZH21B
ZR22G (0.15 m)
ZO21P
ZR802G
ZA8F
ZR40H
ZO21S
2.1

General Specifications

Standard Specifications (Oxygen Analyzer)
n
Measurement Object: Oxygen concentration in combustion exhaust gas and mixed gas (excluding
inflammable gases)
Measurement System: Zirconia system
Measurement Range: 0.01 to 100 vol% O
Output Signal: 4 to 20 mA DC (maximum load resistance 550 Ω)
Setting Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O
range
Display Range: 0 to 100 vol% O
Warming-up Time: Approx. 20 min.
Repeatability: (Excluding the case where the reference gas is by natural convection) ± 0.5% F.S. ;
range from 0 to 5 vol% O
from 0 to 25 vol% O
Linearity: (Excluding standard gas tolerance and the case where the reference gas is by natural
convection) (Use oxygen of known concentration (with in the measuring range) as the zero
and span calibration gases.)
pressure within ± 4.9 kPa
pressure within ± 0.49 kPa
pressure within ± 0.49 kPa
Drift:
(Excluding the first two weeks in use and the case where the reference gas is by natural
convection.)
Both zero and span ± 2% F.S. /month
Response Time: Response of 90% within 5 seconds.
(Measured after gas is introduced from calibration gas inlet and analog output starts
changing.)
Safety, EMC, and RoHS conformity standards of ZR22G, ZR802G
Installation altitude: 2000 m or less
Installation category: (IEC61010); II
Pollution degree: (IEC61010); 2
Measurement category: O (other)
Note
· Installation category, called overvoltage category, specifies impulse withstanding voltage. Category II is for
electrical equipment.
· Pollution degree indicates the degree of existence of solid, liquid, gas or other inclusions which reduce dielectric
strength. Degree 2 is the normal indoor environment.
General use separate type detector (See Section 2.2.1)
Probe protector (See Section 2.2.2)
Dust protector (See Section 2.2.3)
High temperature separate type detector (See Section 2.3.1)
Adapter for High temperature probe (See Section 2.3.2)
Zirconia Oxygen/Humidity Analyzer, Converter (See Section 2.4)
Flow setting unit (See Section 2.5.1)
Automatic calibration unit (See Section 2.5.2)
Standard gas unit (See Section 2.6)
2
2
or more and less than 0 to 25 vol% O
2
or more and up to 0 to 100 vol% O
2
± 1% F.S.; 0 to 5 or more and less than 0 to 25 vol% O
± 3% F.S.; 0 to 25 or more and less than 0 to 50 vol% O
± 5% F.S.; 0 to 50 or more and up to 0 to 100 vol% O
<2. Specifications>
(in 1 vol% O
), or partial
2
2
range ± 1% F.S. ; range
2
range
2
range and sample gas
2
range and sample gas
2
range and sample gas
2
IM 11M12G01-02EN
1sh Edition : Mar. 25, 2021-00
2-1