Default Setting; Simple Cell Resistance Measurement; Mode - YOKOGAWA ZR22G User Manual

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

10.5.5

Default Setting

When the analyzer is delivered, or if data are initialized, the blow back settings are by default, as
shown in Table 10.6.
Table 10.6
Item

Mode

Hold time [mm:ss]
Blowback time [min]
Blowback time [sec]
Interval [day]
Interval [hour]
First start date [YY/MM/DD]
First start time [hh:mm]
10.6

Simple cell resistance measurement

When the cell of the zirconia oximeter deteriorates, the cell resistance increases. This function
simply measures the cell resistance without using calibration gases.
10.6.1
MODE
There are three operation modes of simple cell resistance measurement of this instrument:
"None" without simple cell resistance measurement, "Semi-automatic simple cell resistance
measurement" in which simple cell resistance measurement is started by manipulation from
the tach panel, and "Simple cell resistance measurement" in which simple cell resistance
measurement is automatically performed at setup cycle. Here you can select these modes. The
following restrictions apply to each mode:
When "None" is selected
l
Simple cell resistance measurement is not performed.
When "Semi-automatic" simple cell resistance measurement is selected
l
Semi-automatic simple cell resistance measurement is enabled (measurement is not executed
when the start time of automatic simple cell resistance measurement is reached).
When "Automatic, semi-automatic" simple cell resistance measurement is selected
l
Both automatic and semi-automatic simple cell resistance measurement mode are enabled.
(1) "Converter menu" > "Maintenance" >"Simple cell resist. setting"
(2) When "Measurement mode" is selected in the "Simple cell resist. setting", you can select
"None", "Semi-automatic", "Auto, Semi-automatic."
Figure 10.21
Blow back Default Setting
Default setting
None
10:00
10
0s
30
0
00/01/01
00:00
Simple cell resistance setting