(1) Ambient temperature of the detector: 75°C or less
ZR22G
Detector
CELL(+)
CELL(-)
TC(+)
TC(-)
CJ(+)
CJ(-)
(2) Ambient temperature of the detector: exceeding 75°C
ZR22G
Detector
CELL(+)
CELL(-)
TC(+)
TC(-)
CJ(+)
CJ(-)
Figure 5.8
Wiring for detector output
CAUTION
If shielded cables cannot be used between the detector and the terminal box, for example,
when heat-resistant wiring is used, locate the detector and the terminal box as close together as
possible.
Cable Specifications
Basically, a cable (6-core) that withstand temperatures of at least 80°C is used for this wiring.
When the ambient temperature of the detector exceeds 75°C, install a terminal box, and connect
with the detector using six-piece 600 V silicon rubber insulated glass braided wires.
/CJ Option Specifications
When the /CJ option is specified, connecting the supplied cold junction compensation element
to the CJ (+) and CJ (-) terminals of the converter, and connecting the tip to the FG terminal (see
Fig. 5.9) is not required, and cabling of the CJ terminal is not required.
In addition, the cabling of the TCs is connected using compensating leads for TYPE K.
Shielded cables
Heat-resistant wiring
<5. Wiring>
Terminal box
Shielded cables
IM 11M12G01-02EN
5-8
ZR802G
Converter
CELL(+)
CELL(-)
TC(+)
TC(-)
CJ(+)
CJ(-)
FG
ZR802G
Converter
CELL(+)
CELL(-)
TC(+)
TC(-)
CJ(+)
CJ(-)
FG
1sh Edition : Mar. 25, 2021-00