100
90
80
70
60
50
40
25
30
35
Figure 8.2
Max. and Min. Humidity Set Ranges
"Mixing ratio" setting range
l
The minimum mixing ratio is set to 0 kg/kg or ranges from 0.201 to 0.625 kg/kg. The maximum
"mixing ratio" setting ranges from 0.2 to 1.0 kg/kg, and must be greater than 1.3 times plus 0.187
the mixing ratio set for the minimum.
Setting example 1
If the setting (for a 4 mA current) is 0 kg/kg, you must set the maximum (20 mA) point at more
than 0.2 kg/kg.
Setting example 2
If the setting (for a 4 mA current) is 0.201 kg/kg, you must set the maximum (20 mA) point at more
than 0.449 kg/kg, (0.201 3 1.3 + 0.187 kg/kg). (Numbers after the decimal point are rounded up.)
1
Ranges over which oxygen
0.95
concentrations can be set
0.9
0.85
0.8
0.75
0.7
0.65
0.6
0.55
0.5
0.45
0.4
0.201
0.25
Figure 8.3
Max. and Min. Mixing Ratio Set Ranges
Ranges over which oxygen
concentrations can be set
40
45
50
55
Minimum humidity (for a 4-mA current), % H
0.3
0.4
0.35
Minimum mixing ratio, kg/kg at 4 mA
<8. Detailed Data Setting>
Outside ranges
60
65
70
75
O
2
0.45
0.5
IM 11M12G01-02EN
8-3
80
85
90
95
F8-2E.ai
Outside ranges
0.6
0.55
F8-3E.ai
1sh Edition : Mar. 25, 2021-00
100
0.65