YOKOGAWA ZR22G User Manual page 125

Zirconia oxygen/humidity analyzer
Hide thumbs Also See for ZR22G:
Table of Contents

Advertisement

95
85
75
65
55
45
35
25
15
5
5
15
Figure 8.1
Humidity (amount-of-moisture-content) and relative humidity setting range
l
The minimum humidity is set to 0% H
ranges from 25% to 100% H
the minimum. See an example shown below or the graph Figure 8.2.
Numerical values are also equivalent for relative humidity Check the values by replacing vol %
H
O with % RH
2
Setting example 1
If the minimum setting (for a 4 mA point) is 0% H
more than 25% H
O.
2
Setting example 2
If the minimum setting (for a 4 mA point) is 26% H
at more than 44% H
rounded up.
Ranges over which oxygen
concentrations can be set
25
Minimum oxygen concentration, vol%O
Minimum-Maximum setting range of oxygen concentration
O or ranges from 26 to 100% H
2
O, and must be greater than 0.8 times plus 23 the humidity set for
2
O. (26 x 0.8 + 23 = 44 vol% H
2
<8. Detailed Data Setting>
35
45
(for a minimum current of 4 mA)
2
O, you must set the maximum (20 mA) point at
2
O, you must set the maximum (20 mA) point
2
O). (Numbers after the decimal point are
2
IM 11M12G01-02EN
Outside ranges
55
65
O. The maximum humidity
2
1sh Edition : Mar. 25, 2021-00
8-2
75
F8.0E.ai