4.1
System dimensions
Space requirements
Ambient environment
Compressed air
Electrical power
ÄKTA ready XL Operating Instructions 29281616 AE
TeDIS Valid on Date Printed
Site requirements
Property
Width × Depth × Height (approximate
±10 mm)
System dry weight (approximate)
Make sure to have enough free space on the sides and in front of the system when it
has been installed to allow for service and maintenance.
At least 100 mm free space is needed on the back of the system (where the cabinet
door is located) for ventilation.
The following conditions should be avoided:
•
Sources of heat, such as direct sunlight
•
Strong magnetic or electric fields
•
Vibrations
•
Corrosive gas
•
Dust
•
Temperatures outside recommended operating ranges, see
tions, on page
•
Dry and particle free air
•
Airflow 50 NL/min
•
Supply pressure 6.5 to 10 bar g
Refer to
General specifications, on page 150
ments.
150.
Printed on: 15-09-2023 11:27:35 ***
4.1 Site requirements
Value
1295 × 1215 × 1920 mm
715 kg
Section 9.1 Specifica-
for power, voltage and phase require-
4 Installation
77