Site requirements

Property
Width × Depth × Height (approximate
±10 mm)
System dry weight (approximate)
Make sure to have enough free space on the sides and in front of the system when it
has been installed to allow for service and maintenance.
At least 100 mm free space is needed on the back of the system (where the cabinet
door is located) for ventilation.
The following conditions should be avoided:
Sources of heat, such as direct sunlight
Strong magnetic or electric fields
Vibrations
Corrosive gas
Dust
Temperatures outside recommended operating ranges, see
tions, on page
Dry and particle free air
Airflow 50 NL/min
Supply pressure 6.5 to 10 bar g
Refer to
General specifications, on page 150
ments.
150.
Printed on: 15-09-2023 11:27:35 ***
4.1 Site requirements
Value
1295 × 1215 × 1920 mm
715 kg
Section 9.1 Specifica-
for power, voltage and phase require-
4 Installation