6.2.6 Replacing the pressure gauge

Tools required:
Crosstip screwdriver, size 1,
Open-ended spanner SW 7.
1. Unscrew the base section of the housing (see
steps
to
in chapter "6.2.3 Dismantling
11.
13.
the housing" on page 14).
2. Using an open-ended spanner (SW 7), re-
lease the swivel screw connection
pressure gauge
.
6
3. Press the pressure gauge
mounting
.
14
Tip:
You will find the pressure gauge easier to remove
if you dribble a small amount of spirit between
the pressure gauge and the mounting.
4. Wet the edge of the new pressure gauge
with a little spirit and press it into the
mounting
.
14
Important:
Observe the installation position so that the dis-
play remains clearly legible!
5. Pass the rubber bead over the pressure gauge
with your fingers.
6. Screw the swivel screw connection
pressure gauge
.
6
Take care to ensure the correct installation posi-
tion. The tube must not protrude beyond the hous-
ing.
7. Join the two halves of the housing together
(see "6.2.4 Assembling the housing" on
page 16).
8. Perform a final check (see "3. Final Check" on
page 7).