6.5.2

Treatment Values Monitored by External Equipment

Monitoring of Expiratory CO
For monitoring expiratory CO
patient circuit.
The device shall comply with ISO 80601-2-55.
Monitoring of Oxygen Saturation
For monitoring the patient air oxygen saturation, an external monitoring device shall be
connected to the patient circuit.
The device shall comply with ISO 80601-2-55 and have a high oxygen level alarm.
Doc. 007229 D-3
EPAP (Expiratory Positive Airway Pressure) is the lowest pressure that is recorded dur-
ing the latest expiratory phase.
Leakage
Leakage is the average calculated leak (l/min) over the last breath, with a breath by
breath update.
MVe
Mve (Minute Volume, expiratory) is calculated as the Tidal Volume multiplied with the
Total Breath Rate.
Vte
Vte (Tidal volume, expiratory) is the expired tidal volume for each breath.
% in TgV
Percentage of breaths within the target volume window.
Rise Time
Rise Time displays the duration of the pressure increase during the inspiration phase.
SpO2
SpO
(Saturation of Peripheral Oxygen) displays the patient's oxygen saturation. as meas-
2
ured by the SpO
module.
2
This value is only displayed if the SpO
Displayed ranges: According to the manufacturer's specifications.
Pulse Rate
Pulse Rate displays the patients pulse rate as measured by the SpO2 module.
This value is only displayed if the SpO
Displayed ranges: According to the manufacturer's specifications.
module is connected.
2
module is connected.
2
2
, an external monitoring device shall be connected to the
2
How to Use the Vivo 2
User manual
69