Inspection Equipment And Tools - Breas Clearway 2 Service Manual

Hide thumbs Also See for Clearway 2:
Table of Contents

Advertisement

Inspection Equipment and Tools

The following equipment is required to perform an inspection on the
Clearway 2:
• Gas flow analyser.
or
• Manometer capable of measuring at least ±70 cmH
0.75%) and a flow meter capable of measuring at least 300 sl/m (± 2%).
• 70% Alcohol cleaning wipes
• Test Lung (part no 6011)
• 6mm leak port (part no 00997)
• Clearway Breathing Circuit (part no 0960.101)
• Medical Electrical Safety tester
The following tools are required to open the case, remove and replace
components when necessary:
• Torx screwdriver T15 x 200mm (To open the case)
• Pozidrive screwdriver No. 1
• Hex screwdriver 2mm
• Hex screwdriver 2.5mm (To open the battery cover)
• Soldering iron, to remove and replace the alarm battery every 5 years
9
Maintenance Instructions
NIPPY Clearway 2 Service Manual
O (± 68.6 hPa) (±
2
2210.2657 Rev 3.0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents