Commissioning
[Sensor not dry] continues to be displayed
If
despite the conductivity sensor having been
dried, then you will have to wait some time until
the controller has detected the sensor as dry.
Once you have selected the sensor type, the
prompt automatically appears asking whether
the sensor parameters (zero point) have to be
determined. You can initiate this prompt man‐
ually as follows:
Continuous display ➨ Menu ➨
[Measurement] ➨
[Measuring channel X Conductivity]
[Sensor parameter adjustment] ➨
1.
Use the arrow keys to select
[Automatically determine sensor param‐
eters] .
2.
Continue with
You will see the display showing
ð
[Sensor dry] and
[Automatically determine sensor
parameters] .
3.
Continue with
You will see the display with the
ð
message
[Sensor parameters are automati‐
cally determined] .
The sensor parameters are auto‐
matically carried over.
68
or
or
.
.
.
or

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DULCOMETER diaLog DACb and is the answer not in the manual?

Questions and answers

Natalie
January 25, 2025

How do I calibrate measuring channel

User image 67945bcac35fb
1 comments:
Mr. Anderson
May 14, 2025

To calibrate a measuring channel on a ProMinent diaLog DACb:

1. Select the channel to be calibrated (e.g., Channel 1: Chlorine).
2. Access the [Calibration] menu from the continuous display.
3. Follow the on-screen instructions to perform the calibration.
4. Adjust the value using the appropriate buttons if prompted.
5. Import the calibration result into the controller memory by pressing the confirmation button.
6. The controller will return to the continuous display and use the new calibration values.

If calibration values are outside specified limits, an error message appears, and the calibration is not saved. In that case, check conditions, correct any issues, and repeat the calibration. Each measuring channel must be calibrated separately.