10.2
Calibrating the ORP
Sensor
10.2.1
Selecting the calibration
process for ORP
Selecting the calibration process
There are two calibration processes available
for calibrating the controller:
1-point (with buffer solution)
n
Data input
n
1.
Continuous display ➨
CAL ORP
Offset
Last calibration
CAL setup
Calibration process
Pot. equalisation
continue with <CAL>
Fig. 47: [ORP] calibration menu
The calibration menu is displayed.
ð
2.
Use
to select the Setup menu or start
calibration by pressing
Selecting the calibration process
[CAL Setup] : Press
3.
The menu for selecting the calibra‐
ð
tion process appears.
4.
Using the arrow keys select the required
[Calibration process] and
menu item
press
The input window appears.
ð
5.
Use the arrow keys to select the calibra‐
tion process and press
6.
Continue with
0.0 mV
13:26:11
11/04/2013
1 point
No
A1027
You can now start your chosen cali‐
ð
bration process.
10.2.2
1-point calibration of ORP
sensor (CAL)
Correct sensor operation
Correct measuring and metering is
–
only possible if the sensor is
working perfectly
Refer to the sensor's operating
–
instructions
Remove the sensor from the in-
–
line probe housing and re-fit it for
calibration. To do this, refer to the
operating instructions for your in-
line probe housing
ORP sensor calibration
The ORP sensor cannot be calibrated.
Only an [OFFSET] deviation of magni‐
tude ± 40 mV can be set and thus com‐
pensated. If the ORP sensor deviates
by more than ± 40 mV from the refer‐
ence value, then it must be checked in
accordance with the requirements of
the sensor operating instructions.
Calibration
95