Power Supply Failure; Systematic Localization Of Leakages; Mains Power Supply Failure - Dräger Perseus A500 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Perseus A500:
Table of Contents

Advertisement

Systematic localization of leakages

To find causes of leakages, isolate individual
components from the leakage test.
Component
Measure
Sample line
Remove the sample line and
occlude the Luer-Lock connec-
tion on the Y-piece.
Breathing
Disconnect the breathing hoses.
hoses
Connect the inspiratory port and
expiratory port with a hose that
is known to be without leakages.
Connect the breathing bag
directly to the breathing system.
Vaporizers
Remove the vaporizers.
1 Perform a leakage test, see page 105, and use
the leakage assistant if necessary.
2 Contact service personnel if the leakages
cannot be localized.

Power supply failure

Mains power supply failure

If the mains power supply fails, Perseus
automatically switches to the internal battery. A
fully charged battery will maintain operation for at
least 30 minutes.
Remaining battery charge is displayed on the
status display.
The breathing system heating is deactivated
during battery operation. The peak inspiratory
pressure PIP may be limited, but it will equal at
least 55 hPa (cmH
O).
2
Instructions for use Perseus A500 SW 1.1n
CAUTION
Risk of device malfunction
If mains power fails, devices connected to the
auxiliary power sockets are not supplied from the
internal battery.
Ensure an alternative power supply for
connected devices.
Troubleshooting
155